Year : 2015  |  Volume : 25  |  Issue : 4  |  Page : 365-367
Use of a vascular sheath for introduction of radiofrequency ablation probe during radiofrequency ablation of osteoid osteoma


1 Department of Radiology, Paediatrics Orhtopaedics Unit, Christian Medical College, Vellore, Tamil Nadu, India
2 Department of Orhtopaedics, Paediatrics Orhtopaedics Unit, Christian Medical College, Vellore, Tamil Nadu, India

Click here for correspondence address and email

Date of Web Publication12-Nov-2015
 

   Abstract 


Radiofrequency ablation (RFA) has now become a preferred treatment option for osteoid osteoma. Accurate placement of RFA probe into the nidus of osteoid osteoma is important for good clinical outcome. Various methods and techniques have been described in the literature available. We describe the technique of using a vascular access sheath for introduction of RFA probe after bone drilling, which prevents loss of access to drill track and also serves as a pathway for accurate placement of RFA probe, thereby reducing the risk of damage to the RFA probe tip itself and the surrounding soft tissue.

Keywords: Computed tomography scan; osteoid osteoma; radiofrequency ablation

How to cite this article:
Ahmed M, Keshava SN, Moses V, Mammen S, Jacob KM, Madhuri V. Use of a vascular sheath for introduction of radiofrequency ablation probe during radiofrequency ablation of osteoid osteoma. Indian J Radiol Imaging 2015;25:365-7

How to cite this URL:
Ahmed M, Keshava SN, Moses V, Mammen S, Jacob KM, Madhuri V. Use of a vascular sheath for introduction of radiofrequency ablation probe during radiofrequency ablation of osteoid osteoma. Indian J Radiol Imaging [serial online] 2015 [cited 2019 Oct 15];25:365-7. Available from: http://www.ijri.org/text.asp?2015/25/4/365/169460



   Introduction Top


Osteoid osteoma (OO) is a benign bone tumor mostly occurring in adolescents or young adults.[1] It is radiographically characterized by dense sclerosis surrounding a small radiolucent nidus.[1] Radiofrequency ablation (RFA) has now become a preferred treatment option. Compared to surgery, it is less invasive and morbid and requires shorter hospital stay with earlier weight bearing. RFA causes thermal coagulation necrosis of the tumor nidus.[1],[2] Clinical outcome depends on the complete thermal necrosis of the nidus, which in turn largely depends on the accurate placement of RFA probe into the nidus.[1] Various techniques have been described for the localization of nidus, guidance and tracking for placement of the RFA probe.[1],[2],[3] Many of these techniques have been described using commercially available introduction systems such as RITA StarBurst Hard Tissue Access System (AngioDynamics, Inc., GA, USA).[2],[3] RFA probe is introduced into the nidus after drilling through the cortex of the bone. However, introduction of RFA probe without any introduction system, even on computed tomography (CT) guidance, is often difficult and time consuming due to small diameter of the drill track, and also involves the risk of injury to the surrounding soft tissue or damage to the RFA probe tip itself. Vascular sheath for RFA was earlier used by Mahnken et al. in their two series.[4],[5] They introduced the sheath before drilling to secure the drill track and protect the surrounding soft tissue. Details of the technique of introduction of vascular sheath before drilling were not mentioned in their series. We describe a technique using a vascular sheath for accurately introducing the RFA probe into the nidus after drilling the bony cortex under conventional CT guidance.


   Technique Top


We did a retrospective analysis of 14 cases (12 males and 2 females, age range 6-23 years, mean 14.1 years) where we employed this technique, mostly in deeply located lesions [Table 1]. Biopsy of the nidus was not done and diagnosis of OO was established by characteristic clinical and radiographic features. Complete ablation of the nidus was achieved in all the cases without any recurrence of pain in the follow-up period (range 1-18 months; mean 8.2 months). There was no complication encountered in our small series. In one case, while inserting the RFA probe, there was perforation of the sheath with displacement of needle through the perforation into the soft tissue track. The probe was pulled out and reinserted.
Table 1: List of patients evaluated including location of OO and follow up

Click here to view


We used conventional CT guidance (single slice SOMATOM Emotion; Siemens, Erlangen, Germany), a manual (Aesculap Inc.-B Braun, Center Valley, PA) or battery-operated (Stryker Corp., Kalamazoo, MI) bone drill machine with cannulated drill bit (2.5-4.5 mm) (Zimmmer, Warsaw IN), a commercially available RITA SDE StarBurst probe (17 gauge, 2 cm diameter, 12 cm; Angiodynamics, Inc., GA, USA), and generator (Model 1500X; Angiodynamics, Inc., GA, USA) for RFA. Ablation was for 5 min at a target temperature of 90°C. Technique is described in the following steps.

Step 1. Localization of nidus of OO and drilling

Formal informed consent was obtained before the procedure from the patient or the patient's legal guardian if the patient was a minor. Patient was placed in an appropriate position on the CT scan table after administering general anesthesia. After initial topogram with a radio-opaque marker on the skin surface, limited CT scan of the area of interest was taken with 3 mm slice thickness. Nidus of the OO was localized, and depth, trajectory, and angle for bone drilling from skin surface were assessed on the CT scan workstation. Drilling track through the neurovascular bundle was avoided. Skin surface was marked, prepared with betadine 10% solution, and draped. We took the help of an orthopedic surgeon for bone drilling. Intermittent limited CT scan cuts were taken for guiding the drilling in correct direction and depth.

Step 2. Exchange of drill bit with vascular sheath

After drilling in appropriate trajectory and depth immediately posterior to the nidus, manual or battery-operated drill machine was separated keeping the cannulated drill bit in situ [Figure 1]a. Back and stiff end of a short (80 cm), 0.035-inch Amplatz guidewire (Cook Inc., Bloomington, IN, USA) was introduced through the cannulated drill bit. Drill bit was removed keeping the Amplatz guidewire in situ. Now a 5 or 6 F, short (11 cm) vascular access sheath (AVANTI+; Cordis Corp., FL, USA) (along with its dilator), cut and separated from its hub [Figure 1]b, was introduced into the drilled bony track over the Amplatz guidewire [Figure 1]c.
Figure 1(A-E): (A) Drill bit (small arrow) in the nidus (large arrow) of the OO in lateral condyle of the femur. (B) Vascular access sheath with dilator (arrow) cut and separated from its hub. (C) Vascular access sheath along with dilator (arrow) introduced into the nidus over Amplatz wire (small arrow). (D) RFA probe (large arrow) was introduced through the vascular access sheath (small arrow) into the nidus after removing Amplatz guidewire and dilator. (E) Axial CT image shows RFA probe (large arrow) in the nidus (small arrow) of OO

Click here to view


Step 3. Introduction of RFA probe

Amplatz guidewire and sheath dilator were removed and the RFA probe was introduced without difficulty directly into the nidus through the vascular introduction sheath [Figure 1]d. Limited CT scan cuts were taken to confirm the exact position of the probe tip in the nidus [Figure 1]e. After confirming the position of the probe in the nidus, the sheath was withdrawn 2-3 cm, and in some cases, it was cut longitudinally and peeled away. RFA electrodes were deployed in the nidus and ablation completed by connecting it to generator after placing dispersion electrode pads.

After completion of the ablation, the probe was removed and post procedure CT scan was done. Skin puncture was closed and dressed with adhesive bandage. All the patients were followed up on the same day in the ward and were discharged the next day. They were seen in the OPD within 1 week post procedure. Non-weight-bearing and avoidance of sports or strenuous work on the treatment side were advised for 4-6 weeks. Oral analgesics and prophylactic antibiotics were advised for 3-7 days post procedure.


   Discussion Top


Accurate placement of RFA probe into the nidus of OO is important for good clinical outcome. Various methods and techniques have been described for RFA probe guidance and introduction, including 14 G bone penetration cannula with stylet for introduction of biopsy needle and RFA probe, Bonopty coaxial bone biopsy system (Radi Medical Systems, Uppsala, Sweden), along with dedicated Sluyter-Mehta 20 G thermal ablation cannula (Radionics, Burlington, MA), and RITA StarBurst Hard Tissue Access System.[1],[2],[3],[6],[7] Bonopty bone biopsy system consists of a 14-G and 95-mm-long penetration cannula that allows a 15-G drill and bone biopsy needle and introduction of a 20-G, 145-mm-long RFA cannula and probe through it.[4] RITA StarBurst Hard Tissue Access System also consists of an 11 G cannula and a stylet.[2],[3] RFA probe is introduced through the cannula after removing the stylet. Penetration cannula helps in access to nidus and serves as a pathway for introduction of drill, biopsy needle, and RFA probe.[1],[6] Some authors have used Jamshidi or Ackermann bone biopsy needle to gain access to the nidus, which is again a coaxial system consisting of an outer cannula and a trocar, and the RFA probe is inserted through the outer cannula into the nidus after biopsy.[8] All the above techniques use outer cannula as a coaxial system, which prevent loss of access to the bony track and allows introduction of biopsy needle and RFA probe. Although commercially available introduction systems can be used to introduce the probe precisely into the nidus.[2],[3],[6],[7] these increase the cost of the procedure and are therefore not commonly used.

Mahnken et al.,[4],[5] in their series, placed an introducer vascular sheath before drilling close to the cortical bone to protect the soft tissue and to secure the drill track. Drilling and RFA were done through the sheath. The sheath was placed in the soft tissue and not inserted into the bony track. The sheath was used without cutting and separating the hub. Our technique is different in that we placed the vascular sheath after the bone drilling and inserted it into the bony track. Since the sheath was inserted into the bony track, it was withdrawn before RFA. Hub of the sheath was cut and separated for easy removal of the sheath after the introduction of RFA probe.

In conclusion, use of a vascular sheath during RFA of OO can be a good alternative to other available introduction systems. It secures access to drill track and serves as a pathway for accurate placement of RFA probe, thereby reducing the risk of damage to the RFA probe tip itself and the surrounding tissues.

 
   References Top

1.
Motamedi D, Learch TJ, Ishimitsu DN, Motamedi K, Katz MD, Brien EW, et al. Thermal ablation of osteoid osteoma: Overview and step by-step guide. Radiographics 2009;29:2127-41.  Back to cited text no. 1
    
2.
Busser WM, Hoogeveen YL, Veth RP, Schreuder HW, Balguid A, Renema WK, et al. Percutaneous radiofrequency ablation of osteoid osteomas with use of real-time needle guidance for accurate needle placement: A pilot study. Cardiovasc Intervent Radiol 2011;34:180-3.  Back to cited text no. 2
    
3.
Mylona S, Patsoura S, Galani P, Karapostolakis G, Pomoni A, Thanos L. Osteoid osteomas in common and in technically challenging locations treated with computed tomography-guided percutaneous radiofrequency ablation. Skeletal Radiol 2010;39:443-9.  Back to cited text no. 3
    
4.
Mahnken AH, Bruners P, Delbrück H, Günther RW. Radiofrequency ablation of osteoid osteoma: Initial experience with a new monopolar ablation device. Cardiovasc Intervent Radiol 2011:34;579-84.  Back to cited text no. 4
    
5.
Mahnken AH, Tacke JA, Wildberger JE, Günther RW. Radiofrequency ablation of osteoid osteoma: Initial results with a bipolar ablation device. J Vasc Interv Radiol 2006;17:1465-70.  Back to cited text no. 5
    
6.
Pinto CH, Taminiau AH, Vanderschueren GM, Hogendoorn PC, Bloem JL, Obermann WR. Technical considerations in CT-guided radiofrequency thermal ablation of osteoid osteoma: Tricks of the trade. AJR Am J Roentgenol 2002;179:1633-42.  Back to cited text no. 6
    
7.
Donkol RH, Al-Nammi A, Moghazi K. Efficacy of percutaneous radiofrequency ablation of osteoid osteoma in children. Pediatr Radiol 2008;38:180-5.  Back to cited text no. 7
    
8.
Cantwell CP, Obyrne J, Eustace S. Current trends in treatment of osteoid osteoma with an emphasis on radiofrequency ablation. Eur Radiol 2004;14:607-17.  Back to cited text no. 8
    

Top
Correspondence Address:
Munawwar Ahmed
Department of Radiology, Christian Medical College, Vellore - 632 004, Tamil Nadu
India
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/0971-3026.169460

Rights and Permissions


    Figures

  [Figure 1]
 
 
    Tables

  [Table 1]



 

Top
 
 
  Search
 
   Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles
    Email Alert *
    Add to My List *
* Registration required (free)  


    Abstract
   Introduction
   Technique
   Discussion
    References
    Article Figures
    Article Tables

 Article Access Statistics
    Viewed1693    
    Printed26    
    Emailed0    
    PDF Downloaded126    
    Comments [Add]    

Recommend this journal