INTERVENTIONAL RADIOLOGY  
Year : 2014  |  Volume : 24  |  Issue : 1  |  Page : 80-83
Utility of C-arm CT in overcoming challenges in patients undergoing Transarterial chemoembolization for hepatocellular carcinoma


Department of Radiology, Amrita Institute of Medical Sciences and Research Center, Amrita Lane, Elamakkara P.O. Cochin, Kerala, India

Click here for correspondence address and email

Date of Web Publication16-Apr-2014
 

   Abstract 

Transarterial chemoembolization (TACE) is the well-known treatment for hepatocellular carcinoma. Multiple digital subtraction angiography (DSA) acquisitions in different projections are required to identify difficult arterial feeders. Moreover, the tell-tale tumor blush can be obscured by proximity to lung base, small size of lesion, and breathing artifacts. C-arm CT is a revolutionary advancement in the intervention radiology suite that allows acquisition of data which can be reformatted in multiple planes and volume rendered incorporating both soft tissue and vascular information like multidetector computed tomography (MDCT). These images acquired during the TACE procedure can provide critical inputs for achieving a safe and effective therapy. This case series aims to illustrate the utility of C-arm CT in solving specific problems encountered while performing TACE.

Keywords: Arteriography digital subtraction; C-arm CT; cone beam computed tomography; hepatocellular carcinoma; transarterial chemoembolization

How to cite this article:
Kulkarni C, Sreekumar K P, Prabhu NK, kannan R R, Moorthy S. Utility of C-arm CT in overcoming challenges in patients undergoing Transarterial chemoembolization for hepatocellular carcinoma. Indian J Radiol Imaging 2014;24:80-3

How to cite this URL:
Kulkarni C, Sreekumar K P, Prabhu NK, kannan R R, Moorthy S. Utility of C-arm CT in overcoming challenges in patients undergoing Transarterial chemoembolization for hepatocellular carcinoma. Indian J Radiol Imaging [serial online] 2014 [cited 2018 Jan 23];24:80-3. Available from: http://www.ijri.org/text.asp?2014/24/1/80/130709

   Introduction Top


Transarterial chemoembolization (TACE) is the well-known treatment for hepatocellular carcinoma (HCC). [1],[2] Prior to TACE, multidetector computed tomography (MDCT) can locate the lesion, map the arterial feeders, and detect extrahepatic tumor supply. Accurate superselective positioning of microcatheter ensures optimum drug delivery to the tumor and minimizes the nontarget embolization during TACE. However, tortuous arterial anatomy often mandates multiple digital subtraction angiography (DSA) acquisitions in oblique projections to identify the arterial feeders. Moreover, the tell-tale tumor blush can be obscured by proximity to lung base, small size of the lesion, and breathing artifacts. C-arm CT is a revolutionary advancement in the intervention radiology suite that allows acquisition of data which can be reformatted in multiple planes and volume rendered incorporating both soft tissue and vascular information like MDCT. These images acquired during the TACE procedure can provide critical inputs for achieving a safe and effective therapy.

Aim

This case series aims to illustrate the utility of C-arm CT in solving specific problems encountered while performing TACE.


   Materials and Methods Top


We included five patients (four males and one female) with average age of 58.4 years who underwent TACE for HCC.

Technique of C-arm CT

All procedures were performed using a floor-mounted, single-plane, flat panel DSA unit (Axiom Artis Zee; Siemens, Erlangen, Germany). Using standard femoral arterial access, selective celiac and superior mesenteric artery (SMA) angiograms were done to identify putative tumor feeders. A 2.8 F microcatheter (Progreat) was then placed super selectively into the tumor feeder and TACE performed. However, when the feeder could not be clearly isolated from the adjacent branches or when multiple feeders were supplying tumor or when tumor blush could not be clearly visualized, a C-arm CT was performed. Using a power injector, 8-10 mL contrast with 30%-50% dilution was injected directly into the microcatheter at a flow rate of 2 mL/sec (proper hepatic artery) or 1 mL/sec (right hepatic artery, left hepatic artery). Acquisition delay for arterial imaging was 2-3 sec and for parenchymal imaging was 4-5 sec. [3] Images were acquired during breath-hold with single rotation of C-arm around the patient. Three-dimensional (3D) CT like images were reconstructed and analyzed in dedicated workstation (Leonardo; Siemens, Erlangen, Germany).

Observations

The five cases included depict the use of C-arm CT to overcome various technical difficulties encountered while performing TACE for HCC. The summary of cases is presented in [Table 1].
Table 1: Summary of the cases

Click here to view



   Discussion Top


3D DSA technology was primarily used in neuro-interventions. [4],[5] Liapi et al. were the first to use 3D DSA technology to delineate the vascular anatomy before TACE in two patients. [6] As the detectability of low contrast structures was limited on 3D DSA, the angio-CT system was developed. [7] Initial C-arm CT performed using image intensifier system had limited spatial resolution. [8] Recently, the use of flat panel detectors in C-arm CT has significantly improved the contrast and spatial resolution, providing CT-like images. [9] Depicting both vascular anatomy and soft tissue is beneficial during TACE because the relationship of the tumor with its arterial supply can be clearly identified. [7],[10],[11],[12]

Principles of C-arm CT

The C-arm CT is basically a cone beam CT (CBCT). It allows volumetric data acquisition covering large anatomic area in a single rotation of the source and detector around the stationary patient. The fundamental difference between the conventional MDCT and CBCT is that CBCT acquires information using a single two-dimensional detector (flat panel detector), while MDCT uses multiple one-dimensional detectors. [13] Flat panel detectors enable direct conversion of X-ray energy into digital signal with high resolution. The detector consists of a screen of cesium iodide (CsI) scintillator crystals arranged on a matrix of photoiodides embedded in a solid-state amorphous silicon (aSi: H) or selenium layer. [14] The reconstruction algorithm used in CBCT is a modified Feldkamp algorithm. [15] CBCT with flat panel detector improves the spatial resolution and decreases the radiation dose to the patient. C-arm CT has low contrast resolution as compared to MDCT because of increased scatter of X-rays. Moreover, CsI detectors have lower dynamic range and limited temporal resolution, which further decrease the contrast resolution of C-arm CT. In addition, due to the relatively long acquisition times, the technique is prone for respiratory movement artifacts. [16],[17]

Applications in TACE

Vascular imaging

C-arm CT helps to identify the vascular anatomy, multiple tumor feeders including extrahepatic supply, and tumor blush. For hypervascular lesions seen on DSA, C-arm CT creates an accurate road map to the target lesion. Superselective catheterization can be achieved without resorting to multiple standard DSA views which involve incremental contrast and radiation dose. In case 1, initial DSA showed tumor blush in segment 4a with probable feeder from middle hepatic artery. However, middle hepatic artery injection showed no tumor blush. C-arm CT performed with microcatheter tip in proper hepatic artery revealed the tumor feeder arising from the left hepatic artery. This confusion was due to overlapping arteries on antero-posterior projection [Figure 1]. Accurate road map will prevent inadvertent catheterization of normal arteries and achieves optimal delivery of chemotherapeutic agent to the tumor. C-arm CT can detect feeder arteries not identified on DSA. By identifying multiple feeders, it facilitates complete treatment. Iwazawa et al., in their study of 33 patients showed that C-arm CT is superior (sensitivity 96.9%) to DSA (sensitivity 77.2%) for identifying the tumor feeding arteries during TACE for HCC. [18]
Figure 1 (A, B): A 60-year-old female with segment 4 HCC. Initial digital subtraction angiography (DSA) shows tumor blush (arrow, A). Injection through middle hepatic artery showed no tumor blush. Contrast-enhanced C-arm CT coronal maximum intensity projection (MIP) reformatted image shows enhancement of tumor (arrowheads, B) and tumor feeder from left hepatic artery (arrow, B)

Click here to view


Parenchymal imaging

C-arm CT has the ability to identify parenchymal lesions that are not adequately visualized on DSA. In case 2, the segment 4 lesion was not clearly seen on DSA. C-arm CT showed the tumor blush with feeder from the right hepatic artery and TACE was performed successfully [Figure 2]. In case 3, C-arm CT convincingly showed the tumor blush along the margins of previously ablated segment 8 lesion. This was not seen on DSA [Figure 3]. Thus, hypovascular and small HCC can be confidently targeted using C-arm CT. As C-arm CT gives soft tissue information, the findings of C-arm CT and MDCT or magnetic resonance imaging (MRI) can be easily correlated and this helps both lesion targeting and follow-up comparison [Figure 4]. C-arm CT detects lesions in difficult locations like segments 7 and 8 which can be obscured on DSA due to respiratory movement artifacts [Figure 4]. C-arm CT can also be used to assess the therapeutic endpoint after TACE. [3] Presence of areas in tumor without lipidiol accumulation or residual enhancement indicates additional hepatic or extrahepatic tumor supply. Case 5 had diffuse right lobe HCC with multiple feeders from the right hepatic artery which were selectively entered and TACE performed. At the end of procedure, we performed C-arm CT to assess the response which showed uniform lipidiol uptake in the tumor area. No abnormal blush or additional feeders were seen, thus confirming the completeness of the procedure [Figure 5].
Figure 2 (A, B): A 55-year-old male presented with segment 4 HCC. Contrast-enhanced CT (CECT) axial image shows no significant enhancement of lesion (arrowheads, A). Contrast-enhanced C-arm CT coronal MIP reformatted image shows enhancement of tumor (arrowheads, B) and tumor feeder from right hepatic artery (arrow, B)

Click here to view
Figure 3 (A, B): A 63-year-old male presented with marginal recurrence of previously ablated segment 8 lesion. CECT coronal reformatted image shows arterial enhancement along medial margin of the previously ablated lesion (arrowheads, A). The lesion was not seen on DSA. Contrast-enhanced C-arm CT coronal MIP reformatted image shows the enhancement at recurrence site (arrowheads, B) and tumor feeder (arrow, B)

Click here to view
Figure 4 (A-C): A 68-year-old male with segment 8 HCC. The lesion was not seen on DSA (A). Contrast-enhanced C-arm CT coronal MIP reformatted image shows the tumor (arrowheads, B) and tumor feeder (arrow, B). TACE was performed with lipidiol and adriamycin. One year follow-up CECT coronal reformatted image shows lipidiol uptake in lesion (arrowheads, C)

Click here to view
Figure 5 (A, B): A 46-year-old man with diffuse right lobe hepatocellular carcinoma. DSA shows diffuse tumor blush (arrowheads, A) supplied by multiple tumor feeders from replaced right hepatic artery, which were embolized with lipidiol and adriamycin. Post-TACE C-arm CT coronal MIP reformatted image shows uniform uptake of the lipidiol throughout the tumor area (arrowheads, B)

Click here to view


To conclude, C-arm CT is a useful adjunct technology in intervention radiology suite that assists an interventional radiologist in performing TACE.

 
   References Top

1.Lo CM, Ngan H, Tso WK, Liu CL, Lam CM, Poon RT, et al. Randomized controlled trial of transarterial lipiodol chemoembolization for unresectable hepatocellular carcinoma. Hepatology 2002;35:1164-71.  Back to cited text no. 1
    
2.Llovet JM, Real MI, Montana X, Planas R, Coll S, Aponte J, et al. Arterial embolisation or chemoembolisation versus symptomatic treatment in patients with unresectable Hepatocellular carcinoma: A randomised controlled trial. Lancet 2002;359:1734-9.  Back to cited text no. 2
    
3.Wallace MJ. C-Arm Computed tomography for guiding hepatic vascular interventions. Tech Vasc Interv Radiol 2007;10:79-86.  Back to cited text no. 3
[PUBMED]    
4.Schueler BA, Kallmes DF, Cloft HJ. 3D cerebral angiography: Radiation dose comparison with digital subtraction angiography. Am J Neuroradiol 2005;26:1898-901.  Back to cited text no. 4
    
5.Prestigiacomo CJ, Niimi Y, Setton A, Berenstein A. Three-dimensional rotational spinal angiography in the evaluation and treatment of vascular malformations. Am J Neuroradiol 2003;24:1429-35.  Back to cited text no. 5
    
6.Liapi E, Hong K, Georgiades CS, Geschwind JF. Three-dimensional rotational angiography: Introduction of an adjunctive tool for successful transarterial chemoembolization. J Vasc Interv Radiol 2005;16:1241-5.  Back to cited text no. 6
    
7.Hirota S, Nakao N, Yamamoto S, Kobayashi K, Maeda H, Ishikura R, et al. Cone-beam CT with flat-panel-detector digital angiography system: Early experience in abdominal interventional procedures. Cardiovasc Intervent Radiol 2006;29:1034-8.  Back to cited text no. 7
    
8.Feldkamp LA, Davis LC, Kress JW. Practical cone-beam algorithm. J Opt Soc Am 1984;1:612-9.  Back to cited text no. 8
    
9.Ning R, Chen B, Yu R, Conover D, Tang X, Ning Y. Flat panel detector based cone-beam volume CT angiography imaging: System evaluation. IEEE Trans Med Imaging 2000;19:949-63.  Back to cited text no. 9
    
10.Meyer BC, Frericks BB, Albrecht T, Wolf KJ, Wacker FK. Contrast-enhanced abdominal angiographic CT for intra-abdominal tumor embolization: A new tool for vessel and soft tissue visualization. Cardiovasc Intervent Radiol 2007;30:743-9.  Back to cited text no. 10
    
11.Wallace MJ, Murthy R, Kamat P, Gupta S, Hicks ME, Ahrar K, et al. C-arm CT: Oncologic experience with hepatic arterial interventions. Chicago: Ragiological Society of North America; 2006.  Back to cited text no. 11
    
12.Liu DM, Salem R, Bui JT, Courtney A, Barakat O, Sergie Z, et al. Angiographic considerations in patients undergoing liver-directed therapy. J Vasc Interv Radiol 2005;16:911-35.  Back to cited text no. 12
    
13.Orth RC, Wallace MJ, Kuo MD; Technology Assessment Committee of the Society of Interventional Radiology. C-arm C. C-arm Cone-beam CT: General Principles and technical considerations for use in interventional radiology. J Vasc Interv Radiol 2008;19:814-20.  Back to cited text no. 13
    
14.Baba R, Konno Y, Ueda K, Ikeda S. Comparison of flat-panel detector and image intensifier detector for cone-beam CT. Comput Med Imaging Graph 2002;26:153-8.  Back to cited text no. 14
    
15.Gupta R, Grasruck M, Suess C, Bartling SH, Schmidt B, Stierstorfer K, et al. Ultra-high resolution flat-panel volume CT: Fundamental principles, design architecture, and system characterization. Eur Radiol 2006;16:1191-205.  Back to cited text no. 15
    
16.Siewerdsen JH, Jaffray DA. Cone beam computed tomography with a flat-panel imager: Magnitude and effects of x-ray scatter. Med Phys 2001;28:220-31.  Back to cited text no. 16
    
17.Siewerdsen JH, Moseley DJ, Bakhtiar B, Richard S, Jaffray DA. The influence of antiscatter grids on soft-tissue detectability in cone-beam computed tomography with flat-panel detectors. Med Phys 2004;31:3506-20.  Back to cited text no. 17
    
18.Iwazawa J, Ohue S, Mitani T, Abe H, Hashimoto N, Hamuro M, et al. Identifying feeding rteries during TACE of Hepatic Tumors: Comparison of C-Arm CT and digital subtraction angiography. AJR Am J Roentgenol 2009;192:1057-63.  Back to cited text no. 18
    

Top
Correspondence Address:
Chinmay Kulkarni
Department of Radiology, Amrita Institute of Medical Sciences and Research Center. Amrita Lane, Elamakkara P.O., Kochi, Kerala
India
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/0971-3026.130709

Rights and Permissions


    Figures

  [Figure 1], [Figure 2], [Figure 3], [Figure 4], [Figure 5]
 
 
    Tables

  [Table 1]

This article has been cited by
1 Time-Resolved Perfusion Imaging at the Angiography Suite
Julian H.W. Jürgens,Nadine Schulz,Christian Wybranski,Max Seidensticker,Sebastian Streit,Jan Brauner,Walter A. Wohlgemuth,Yu Deuerling-Zheng,Jens Ricke,Oliver Dudeck
Investigative Radiology. 2015; 50(2): 108
[Pubmed] | [DOI]



 

Top
 
 
  Search
 
   Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles
    Email Alert *
    Add to My List *
* Registration required (free)  


    Abstract
   Introduction
    Materials and Me...
   Discussion
    References
    Article Figures
    Article Tables

 Article Access Statistics
    Viewed2125    
    Printed34    
    Emailed0    
    PDF Downloaded197    
    Comments [Add]    
    Cited by others 1    

Recommend this journal