INTERVENTIONAL RADIOLOGY  
Year : 2011  |  Volume : 21  |  Issue : 4  |  Page : 291-293
Measurement of hepatic venous pressure gradient revisited: Catheter wedge vs balloon wedge techniques


1 Department of Radiology, Christian Medical College, Vellore, Tamil Nadu, India
2 Department of Gastroenterology, Christian Medical College, Vellore, Tamil Nadu, India

Click here for correspondence address and email

Date of Web Publication8-Dec-2011
 

   Abstract 

Aims: To evaluate the accuracy of measurement of hepatic venous pressure gradient by catheter wedge as compared to balloon wedge (the gold standard). Materials and Methods: Forty-five patients having a clinical diagnosis of intrahepatic portal hypertension were subjected to the two different types of pressure measurements (catheter wedge and balloon wedge) during transjugular liver biopsy under fluoroscopic guidance. Statistical Analysis: Spearman's rank correlation coefficient, Bland-Altman plot for agreement, and single measure intraclass correlation were used for analysis of data. Results: There was a close correlation between the results obtained by both the techniques, with highly significant concordance (P < 0.0001). Hepatic venous pressure gradients as measured by the catheter wedge technique were either equal to or less than those obtained by the balloon wedge technique.
Conclusions: The difference in hepatic venous pressure gradients measured by the two techniques is insignificant.

Keywords: Balloon wedge hepatic venous pressure; catheter wedge hepatic venous pressure; free hepatic venous pressure; hepatic venous pressure gradient

How to cite this article:
Chelliah S T, Keshava SN, Moses V, Surendrababu NR, Zachariah UG, Eapen C E. Measurement of hepatic venous pressure gradient revisited: Catheter wedge vs balloon wedge techniques. Indian J Radiol Imaging 2011;21:291-3

How to cite this URL:
Chelliah S T, Keshava SN, Moses V, Surendrababu NR, Zachariah UG, Eapen C E. Measurement of hepatic venous pressure gradient revisited: Catheter wedge vs balloon wedge techniques. Indian J Radiol Imaging [serial online] 2011 [cited 2019 Nov 18];21:291-3. Available from: http://www.ijri.org/text.asp?2011/21/4/291/90693

   Introduction Top


Hepatic venous wedge pressure (HVWP) is a dynamic variable reflecting the hepatic sinusoidal pressure and therefore, indirectly, the portal pressure. [1] It is used to calculate the hepatic venous pressure gradient (HVPG), [2] the normal value being between 1 and 5 mmHg. [3],[4] The current gold standard technique for measuring HVWP is the balloon wedge method. [3],[4] The clinical importance of measuring HVPG in portal hypertension is well documented in the literature. [3],[4]

Because of the technical difficulties associated with the balloon wedge technique, the potential for hepatic venous intimal injury, and the added cost of a balloon catheter, an alternative method with equivalent accuracy would be useful. We conducted the current study with the aim of determining whether a significant difference existed in the HVPG if measured using the conventional catheter wedge method.


   Materials and Methods Top


From May 2007 to June 2008, 45 patients having a clinical diagnosis of intrahepatic portal hypertension due to etiologies that included viral (hepatitis B and C), autoimmune, metabolic (Wilson disease), cryptogenic cirrhosis, and noncirrhotic intrahepatic portal hypertension referred to our department for hepatic vein pressure measurements and transjugular liver biopsy (TJLB), were included in this study. Of the 45 patients with a mean age 41.6 ± 12.8 years, 30 were male. Cases of  Budd-Chiari syndrome More Details were excluded.

Definitions

  • HVWP is a dynamic variable indirectly reflecting the sinusoidal pressure. [5]
  • HVPG is the difference in the HVWP and the free hepatic venous pressure (FHVP). [5]


Measurement of hepatic vein pressures

The procedure was done under local anesthesia. The right internal jugular vein was accessed under ultrasonography guidance. A 9F sheath (Cook Medical, Bloomington, USA) was placed instead of a 6F sheath (Cook Medical) if the patient had to undergo a TJLB. The right hepatic vein was selectively cannulated under fluoroscopy using a 5F multipurpose catheter and a 0.035' guidewire (Cordis, Miami, FL, USA). An angiogram was performed to confirm the location of the catheter. Right atrial pressure and FHVP were then recorded. Subsequently, the catheter was wedged against the hepatic parenchyma and three pressures were recorded at different sites . We ensured that there was no contrast in the catheter during pressure measurement. All pressures were measured using an Agilent; monitor with a quartz pressure transducer.

The catheter was exchanged for a balloon catheter (6F/7F Swan Ganz, Edwards Lifesciences, Irvine, USA) over a 0.030'/0.035' guidewire (Emerald, Cordis, Miami, FL, USA). The balloon was inflated using 50% iodinated contrast (iohexol, 300 mgI/mL, Omnipaque, GE Healthcare, Shanghai, China) in normal saline until the balloon flattened against the walls of the hepatic vein and the pressure through the lumen of the balloon catheter was recorded [Figure 1].
Figure 1: (A, B) Contrast study shows hepatic venogram (A) and balloon wedging (B)

Click here to view
Figure 2: Hepatic venous pressure gradient measured using catheter wedge method and balloon wedge method with the line of equality

Click here to view


TJLB was done whenever requested. The sheath was removed and external compression was applied at the puncture site for hemostasis.

Statistical analysis

For statistical analysis of the results we used three different methods. In the first method, we calculated the Spearman's rank correlation coefficient, that is, Spearman's rho (ρ).[6] The second method was the Bland-Altman plot for agreement, [7],[8] which can be used to compare a new measurement technique or method with a gold standard. In the final method of comparison, the correlation coefficient was calculated by single-measure intraclass correlation. [9]


   Results Top


The measured values of both catheter wedge and balloon wedge HVPGs are plotted in [Figure 2], along with the line of equality. Statistical analysis was done using the 3 different methods as explained:{Figure 2}

  1. Spearman's correlation coefficient (ρ) was found to be 0.876, with P < 0.01, showing that there is a significant correlation.
  2. The Bland-Altman plot for agreement between catheter wedge and balloon wedge pressure gradients [Figure 3] showed good correlation.
    Figure 3: Bland-Altman plot for agreement of catheter wedge and balloon wedge techniques

    Click here to view
  3. The intraclass correlation coefficient was 0.8895, with P < 0.001, showing that the correlation is significant.


Subgroup analysis (1-12, 13-20, and >20 mmHg) was not possible because there weren't enough cases in the 2 nd and 3 rd subgroups to allow any significant statistical correlation.

All the patients tolerated the procedure well and there were no postprocedural complications.


   Discussion Top


HVWP measurement plays an integral role in the evaluation of patients with portal hypertension. A few important applications include the assessment of portal pressure in cirrhosis to predict risk of formation of varices (HVPG > 12 mmHg) or the risk of rebleeding from varices (HVPG > 20 mmHg), to assess the response to pharmacologic agents during treatment of portal hypertension (>20% fall in HVPG from baseline), and to help differentiate between presinusoidal (as in the case of noncirrhotic portal hypertension) and sinusoidal causes of portal hypertension. [2],[10],[11] Hence HVPG monitoring provides important information for patients undergoing treatment for the prevention of variceal bleeding or rebleeding, allowing identification of patients who will be effectively protected against the risk of bleeding and those who, by not achieving an HVPG reduction by 20% of baseline or to ≤2 mmHg, remain at a very high risk of bleeding. [12]

There is only one similar study in the literature; that study, done by Barth and Udoffin 1980, compared the HVPGs in a smaller number (11) of patients with alcoholic cirrhosis as measured by balloon wedge and catheter wedge techniques. Variations in results between the 2 methods ranged up to 3 mmHg, with all patients having elevated hepatic wedge pressures of 15-25 mgHg. They found that the difference was insignificant. [1] Our study includes patients of varied etiology and a wide range of HVPGs of 1-30 mmHg.

Comparison of techniques

The catheter wedge technique is relatively easy to perform but requires multiple measurements of HVWP. Sometimes it needs injection of contrast media to confirm the wedge position. There are certain risks associated with this technique, such as perforation of the vessel by the guide wire or catheter or hepatic parenchymal damage by forceful contrast injection. On the other hand, the balloon wedge technique results are more representative of average parenchymal pressure than the standard wedged catheter measurements. It requires a balloon and an exchange wire, is more time consuming, and exposes patients to slightly higher levels of radiation. Enlargement of the puncture site with severe bleeding may occur with dislodgement of the incompletely deflated balloon. So prior to withdrawal of the catheter through the puncture site, complete balloon deflation is necessary. [1] Potential intimal injury is a rare but important complication with the balloon wedge technique.

The total cost of HVPG measurement at our center using the catheter wedge technique is INR 8000. The use of the balloon catheter would result in an additional cost of INR 2000.


   Acknowledgment Top


We acknowledge our sincere thanks to our statistician Mrs. Visalakshi Jeyaseelan, MSc, Department of Biostatistics, Christian Medical College, Vellore, for her guidance and help.

 
   References Top

1.Barth KH, Udoff EJ. Transfemoral balloon catheterization for hepatic wedge pressure measurements. Radiology 1980;135:779-80.  Back to cited text no. 1
[PUBMED]  [FULLTEXT]  
2.Jaime Bosch, Gennaro D'Amico. Portal Hypertension and non surgical management. Schiff's disease of liver. 10 th ed. Vol. 1. Philadelphia, USA: Lippincott Williams & Wilkins; 2007. p. 436-9  Back to cited text no. 2
    
3.Groszmann r. Measurement of portal pressure: When, How and Why to Do It? Clin Liver Dis 2006;10:499-512.   Back to cited text no. 3
    
4.Groszmann RJ, Wongcharatrawee S. The hepatic venous pressure gradient: Anything worth doing should be done right. Hepatology 2004;39:280-2.  Back to cited text no. 4
    
5.Boyer TD. WHVP: Ready for prime Time. Hepatology 2006;43:405-6.  Back to cited text no. 5
    
6.Spearman C. The proof and measurement of association between two things. Am J Psychol 1987;100:441-71.  Back to cited text no. 6
    
7.Bland JM, Altman DG. Measuring agreement in method comparison studies. Stat Methods Med Res 1999;8:135-60.  Back to cited text no. 7
    
8.Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1986;1:307-10.  Back to cited text no. 8
    
9.Fisher RA. Statistical methods for research workers. 12 th edition. New York: Hafner publishing company; 1954. p. 211-47.  Back to cited text no. 9
    
10.Thomas D, Boyer J. Michael Henderson. Portal Hypertension and bleeding esophageal varices, Hepatology. 4 th ed. Vol. 1. Philadelphia, USA: Saunders; 2003. p. 581-630.  Back to cited text no. 10
    
11.Madhu K, Avinash B, Ramakrishna B, Eapen CE, Shyamkumar NK, Zachariah U, et al. Idiopathic non-cirrhotic intrahepatic portal hypertension: Common cause of cryptogenic intrahepatic portal hypertension in a Southern Indian tertiary hospital. Indian J Gastroenterol 2009;28:83-7.  Back to cited text no. 11
    
12.D'Amico G, Garcia-Pagan JC, Luca A, Bosch J. Hepatic vein pressure gradient reduction and prevention of variceal bleeding in cirrhosis: A systematic review. Gastroenterology 2006;131:1611-24.  Back to cited text no. 12
    

Top
Correspondence Address:
S Timothy Chelliah
Department of Radiology, CMCH, Vellore - 632 004, Tamil Nadu
India
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/0971-3026.90693

Rights and Permissions


    Figures

  [Figure 1], [Figure 2], [Figure 3]

This article has been cited by
1 Prediction of portal pressure from intraoperative ultrasonography
Atsushi Nanashima,Takafumi Abo,Junichi Arai,Tetsuro Tominaga,Katsunori Takagi,Koji Mochinaga,Katsuro Furukawa,Takeshi Nagayasu
Journal of Surgical Research. 2014;
[Pubmed] | [DOI]
2 Comprehensive Predictors of Portal Pressure from Functional Liver Reserve in Patients Who Underwent Hepatectomy
Atsushi Nanashima,Takafumi Abo,Junichi Arai,Tomoshi Tsuchiya,Takuro Miyazaki,Katsunori Takagi,Xiaohui Chen,Takeshi Nagayasu
Indian Journal of Surgery. 2014;
[Pubmed] | [DOI]
3 Use of serum vitamin B12 level as a marker to differentiate idiopathic noncirrhotic intrahepatic portal hypertension from cryptogenic cirrhosis
Goel, A. and Ramakrishna, B. and Muliyil, J. and Madhu, K. and Sajith, K.G. and Zachariah, U. and Ramachandran, J. and Keshava, S.N. and Selvakumar, R. and Chandy, G.M. and Elias, E. and Eapen, C.E.
Digestive Diseases and Sciences. 2013; 58(1): 179-187
[Pubmed]
4 Use of Serum Vitamin B12 Level as a Marker to Differentiate Idiopathic Noncirrhotic Intrahepatic Portal Hypertension from Cryptogenic Cirrhosis
Ashish Goel,Banumathi Ramakrishna,Jayaprakash Muliyil,Kadiyala Madhu,K. G. Sajith,Uday Zachariah,Jeyamani Ramachandran,Shyamkumar N. Keshava,R. Selvakumar,George M. Chandy,Elwyn Elias,C. E. Eapen
Digestive Diseases and Sciences. 2012;
[Pubmed] | [DOI]



 

Top
 
 
  Search
 
   Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles
    Email Alert *
    Add to My List *
* Registration required (free)  


    Abstract
   Introduction
    Materials and Me...
   Results
   Discussion
   Acknowledgment
    References
    Article Figures

 Article Access Statistics
    Viewed7618    
    Printed140    
    Emailed1    
    PDF Downloaded330    
    Comments [Add]    
    Cited by others 4    

Recommend this journal