Technical note: CT-guided biopsy of lung masses using an automated guiding apparatus


Barnard Institute of Radiology, Madras Medical College, Government General Hospital, Chennai- 600 003, India

Click here for correspondence address and email

Date of Web Publication4-Aug-2009
 

   Abstract 

Automated guiding apparatuses for CT-guided biopsies are now available. We report our experience with an indigenous system to guide lung biopsies. This system gave results similar to those with the manual technique. Automated planning also appears to be technically easier, it requires fewer number of needle passes, consumes less time, and requires fewer number of check scans.

Keywords: Automated / manual planning; CT-guided needle lung biopsy

How to cite this article:
Chellathurai A, Kanhirat S, Chokkappan K, Swaminathan TS, Kulasekaran N. Technical note: CT-guided biopsy of lung masses using an automated guiding apparatus. Indian J Radiol Imaging 2009;19:206-7

How to cite this URL:
Chellathurai A, Kanhirat S, Chokkappan K, Swaminathan TS, Kulasekaran N. Technical note: CT-guided biopsy of lung masses using an automated guiding apparatus. Indian J Radiol Imaging [serial online] 2009 [cited 2019 Aug 24];19:206-7. Available from: http://www.ijri.org/text.asp?2009/19/3/206/54883

   Introduction Top


CT-guided lung biopsy is a usually done manually, using a standard technique. For some years now, automated systems have been available to guide biopsies. [1],[2] We discuss our experience with a newly developed indigenous system.


   Technique Top


We used PIGA-CT (a robotic five-axes guide arm and planning console) designed by Perfint Healthcare Pvt. Ltd. (Chennai, India); it is an automated apparatus that calculates coordinates on DICOM images from a CT scanner and guides the placement of a needle accurately within the body after insertion [Figure 1].

The apparatus consists of an electromechanical guide arm that provides five degrees of freedom, a computer console for receiving CT images and calculating coordinates, and an RS232 interface for data communication between the guide arm and the computer console. Precise 'point of insertion' and 'point of target' are determined from the images and marked. The apparatus is able to accurately position itself by using the movement in five axes. The manipulator aligns the needle guide. The needle is required to enter the body at the 'point of insertion' and to touch the target at the 'point of target' [Figure 2] and [Figure 3].

Of 36 consecutive CT-guided needle biopsies of the chest performed at our institute between 30 th June 2007 and 24 th January 2008, 18 (group I) were performed using manual planning and 18 (group II) with the automated biopsy system. A four-slice CT scanner (Toshiba; DICOM compatible) was used to localize the lesion and to guide needle placement. No cytopathologists were present at the time of biopsy.

Seven (19%) small post-biopsy pneumothoraces complicated the 36 procedures; none of them required placement of a chest tube. Four cases occurred while using the manual method and three during the automated method. In this study, the technical success was 100% with both the methods. Using the manual method, 11 biopsies (61.1%) yielded sufficient tissue for pathologic evaluation whereas, with the automated apparatus, 12 (66.7%) biopsies gave a definitive diagnosis.


   Discussion Top


We were able to show that the automated system works well and could provide technical and diagnostic success rates similar to those obtained with the manual method. Also, we found that the automated device decreased the number of needle position adjustments and thereby minimized the procedure time. There was no significant difference in the incidence of complications with the two methods.

In our opinion, such automated systems can be extremely useful when the radiologist doing the biopsy has limited experience or when the lesion is situated in difficult locations. Larger trials are required to assess the usefulness and cost-effectiveness of such automated systems in different clinical environments.


   Acknowledgment Top


Dr. Chithrakala Sukumar, PG trainee in MD (pathology), Madras Medical College, Chennai for helping in data collection from pathology department) and Mrs. Basilea Watson, Biostatistician,Madras Medical College, Chennai for helping in statistical analysis of data.

 
   References Top

1.Moulton JS, Moore PT. Coaxial percutaneous biopsy technique with automated biopsy devices: Value in improving accuracy and negative predictive value. Radiology 1993;186:515-22.  Back to cited text no. 1  [PUBMED]  [FULLTEXT]
2.Burbank F, Kaye K, Belville J, Ekuan J, Blumenfeld M. Image-guided automated core biopsies of the breast, chest, abdomen and pelvis. Radiology 1994;191:165-71.  Back to cited text no. 2  [PUBMED]  [FULLTEXT]

Top
Correspondence Address:
Amarnath Chellathurai
52, Khanabagh Street, Triplicane, Chennai- 600 005
India
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/0971-3026.54883

Rights and Permissions


    Figures

  [Figure 1], [Figure 2], [Figure 3]

This article has been cited by
1 The common complications in percutaneous lung biopsy: An analysis of the relevant risk factors
Chen, W.-H., Shen, X.-W., Sun, X.-G., Wang, Z., Zhang, Z.-Y.
Journal of Interventional Radiology. 2012; 21(2): 168-171
[Pubmed]
2 CT-guided biopsies: Quality, complications and impact on treatment - A retrospective initial quality control
De Bucourt, M., Busse, R., Zada, O., Kaschke, H., Weiss, A., Teichgräber, U., Rogalla, P., Hein, P.A.
RoFo Fortschritte auf dem Gebiet der Rontgenstrahlen und der Bildgebenden Verfahren. 2011; 183(9): 842-848
[Pubmed]
3 Interventions and turfs
Jankharia, B.
Indian Journal of Radiology and Imaging. 2009; 19(3): 175
[Pubmed]



 

Top
 
  Search
 
    Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles
    Email Alert *
    Add to My List *
* Registration required (free)  


    Abstract
    Introduction
    Technique
    Discussion
    Acknowledgment
    References
    Article Figures

 Article Access Statistics
    Viewed2700    
    Printed132    
    Emailed0    
    PDF Downloaded539    
    Comments [Add]    
    Cited by others 3    

Recommend this journal