Year : 2009  |  Volume : 19  |  Issue : 3  |  Page : 203-205
Technical note: A simple and effective CO 2 delivery system for angiography using a blood bag


Department of Radiology, Kovai Medical Center and Hospital, Avinashi Road, Coimbatore, Tamil Nadu, India

Click here for correspondence address and email

Date of Web Publication4-Aug-2009
 

   Abstract 

Several angiographic techniques have been developed to image the arterial system, the commonest using iodinated contrast media. Useful as they may be, they are not without disadvantages. One other modality is angiography using CO 2 . Although CO 2 can be used as an alternative contrast medium, delivery systems are expensive to procure. We describe an indigenous and effective delivery system developed at our institute.

Keywords: Carbon dioxide, CO 2 , contrast, DSA

How to cite this article:
Cherian MP, Mehta P, Gupta P, Kalyanpur TM, Jayesh S R, Rupa R. Technical note: A simple and effective CO 2 delivery system for angiography using a blood bag. Indian J Radiol Imaging 2009;19:203-5

How to cite this URL:
Cherian MP, Mehta P, Gupta P, Kalyanpur TM, Jayesh S R, Rupa R. Technical note: A simple and effective CO 2 delivery system for angiography using a blood bag. Indian J Radiol Imaging [serial online] 2009 [cited 2019 Dec 8];19:203-5. Available from: http://www.ijri.org/text.asp?2009/19/3/203/54881

   Introduction Top


Vascular diseases of the arterial system have long been a major cause of morbidity and mortality. Although several noninvasive techniques have evolved in recent times to image the arterial system, the gold standard remains the catheter digital subtraction angiography (DSA). Though the iodinated contrast media used today are relatively safe, they should, however, be used with caution in patients with compromised renal function and in diabetics. Patients with inadequate cardiac function are also poor candidates for angiography with iodinated contrast. Carbon dioxide (CO 2 ) has a role in this subset of patients as an effective and safe contrast agent. [1]

Though gadolinium may be used as a substitute contrast agent [1] in patients with chronic renal failure, there is a chance of developing a potentially fatal complication called nephrogenic systemic fibrosis. [2],[3]

CO 2 has received immense attention in recent times for its excellent safety profile. There is a paucity of studies, related to procedures using CO 2 in the Indian medical literature; a PubMed search in fact yielded no study from India regarding the use of CO 2 in vascular procedures; this is probably due to the fact that dedicated delivery systems for CO 2 are expensive to procure and maintain. This problem has been solved by developing an indigenous injection system at our institute.


   Materials and Methods Top


The system developed at our institute has been used for CO 2 angiograms for many years now. It consists of a standard blood bag (readily available) connected to a CO 2 cylinder [Figure 1]; this is done in order to avoid direct transmission of pressure from the cylinder to the vasculature under examination. The blood bag is then connected to a three-way stopcock [Figure 2], to which are connected a 50-cc plastic syringe and the angiographic catheter to be used [Figure 3].

The blood bag is first filled with CO 2 and the knob on the cylinder is closed. The bag now acts as a CO 2 reservoir. CO 2 is then allowed to freely flow for some time through the three ways in a basin containing sterile saline. This ensures that the tubings are free of air. The blood bag is intermittently filled to ensure that there is adequate CO 2 in this low pressure reservoir. With the CO 2 flowing through the ports of the three ways, a luer lock syringe is connected ensuring that the syring is kept upside down. A 50 cc syringe is used if the injection has to be made in the renal or infra renal aorta. We would use a 20 cc syringe in the iliac and femoral arteries. Seventy five percent of the syringe is filled with CO 2 by gentle aspiration. The catheter is flushed with about 3 cc of CO 2 to prevent explosive delivery of the agent. According to Caridi et al , [1] this explosive delivery is a major cause of patient discomfort. Once the catheter has been flushed, fifty percent of the CO 2 in the syringe is injected to adequately visualize the vessel. The CO 2 displaces blood in the vessel and provides 'negative contrast' [Figure 4]. The image quality can be enhanced by using 'stacking software'. Stacking software is used because the CO 2 column undergoes fragmentation as it travels distally; the software post-processes the images by stacking them one over the other, thus leading to an image in which the entire CO 2 column is visualized (minimum intensity stacking software). This injection system was employed in a prospective study between September 2005 and August 2007, wherein 44 patients underwent lower limb arterial angiograms with CO 2 and iodinated contrast (IC).


   Results Top


Of the 44 cases evaluated with this system, we obtained diagnostic quality angiograms in 42 (95.5%) cases. The system was easy to use and required minimal manpower and expertise to maintain and setup. There was no system malfunction at any time and no complications were encountered.


   Discussion Top


The commercially available delivery systems are not readily available in India and are difficult and expensive to procure. Hawkins et al , [4] developed a plastic bag system in 1995, which they later modified [5] to include O-rings to reduce the possibility of a CO 2 leak; this system is commercially available as the Angioflush III. Our system has been indigenously developed and uses commonly available materials that are cheap to procure and maintain. The use of the blood bag as the reservoir avoids transmission of the high pressure of the gas in the cylinder to the vasculature; it also allows us to fill the syringe with a limited quantity of CO 2 . We circumvented the problem of CO 2 leaks by using a three-way stopcock to which the catheter, syringe and the blood bag are attached. With proper operation of the stopcock, room air contamination can be completely avoided. This is an effective and cheap alternative to the commercially available CO 2 delivery systems.

 
   References Top

1.Caridi JG, Hawkins IF Jr, Klioze SD, LeVeen RF. Carbon dioxide digital subtraction angiography: The practical approach. Tech Vasc Interv Radiol 2001;4:57-65.  Back to cited text no. 1  [PUBMED]  [FULLTEXT]
2.Thomsen HS, Morcos SK, Dawson P. Is there a causal relation between the administration of gadolinium based contrast media and the development of nephrogenic systemic fibrosis (NSF)? Clin Radiol 2006;61:905-6.  Back to cited text no. 2  [PUBMED]  [FULLTEXT]
3.High WA, Ayers RA, Chandler J, Zito G, Cowper SE. Gadolinium is detectable within the tissue of patients with nephrogenic systemic fibrosis. J Am Acad Dermatol 2007;56:21-6.   Back to cited text no. 3  [PUBMED]  [FULLTEXT]
4.Hawkins IF, Caridi JG, Kerns SR. Plastic bag delivery system for hand injection of carbon dioxide. AJR Am J Roentgenol 1995;165:1487-9.  Back to cited text no. 4    
5.Hawkins IF, Caridi JG, Klioze SD, Mladinich CR. Modified plastic bag system with O-ring fitting connection for carbon dioxide angiography. AJR Am J Roentgenol 2001;176:229-32.  Back to cited text no. 5    

Top
Correspondence Address:
Mathew P Cherian
Department of Radiology, Kovai Medical Center and Hospital, Avinashi Road, Coimbatore- 641 014, Tamil Nadu
India
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/0971-3026.54881

Rights and Permissions


    Figures

  [Figure 1], [Figure 2], [Figure 3], [Figure 4]



 

Top
 
  Search
 
    Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles
    Email Alert *
    Add to My List *
* Registration required (free)  


    Abstract
    Introduction
    Materials and Me...
    Results
    Discussion
    References
    Article Figures

 Article Access Statistics
    Viewed2580    
    Printed144    
    Emailed1    
    PDF Downloaded449    
    Comments [Add]    

Recommend this journal